JIUJIUXIN emergency/transport/portable ventilator has three gas source modes:
1. Standard 2L oxygen cylinder;
2. Other oxygen cylinders of various specifications;
3. Oxygen supply on the wall center;
Alarm phenomenon of JIUJIUXIN emergency/portable/transport ventilator: the "SL" letter or pressure gauge symbol flashes on the screen, and the ventilator gives an audible and visual alarm;
Alarm value: when the gas source pressure is lower than 0.28MPa;
Alarm reason: the output pressure of gas source is too low (0.4MPa recommended); the output pressure of pressure reducing valve or central oxygen supply is not stable;
ADD:No.168, Fenghuang West Road, Taizhou, Jiangsu Province, China.
TEL:+86 523-86880973-ext
FAX:+86 523-86196046
QQ :2256378460