Alarm types of emergency ventilator JOGGER

Release:jiuxin-medical Browse:1091

Alarm types of emergency ventilator JOGGER:


Alarms

Alarm Level

Alarm Performances

Alarm Detection Principles

Others

High Airway Pressure Limit Alarm

High Level

Auditory, Symbol Alarm

When the airway pressure is higher than the preset alarm data.


Low Airway Pressure Limit Alarm

Medium Level

Auditory, Symbol Alarm

When the airway pressure is lower than the preset alarm data.

Low Battery Voltage Alarm

Medium Level

Auditory, Symbol Alarm

When the internal battery is undervoltage.

Low Oxygen Source Alarm

High Level

Auditory, Symbol Alarm

When the pressure of oxygen source is lower than 0.28MPa

High Respiratory Rate Alarm

Medium Level

Auditory, Symbol Alarm

When the respiratory rate is higher than the preset alarm data